Purpose:
To evaluate the repeatability of a multispectral-based refractor in central and peripheral refraction measurement, and to assess the agreement of such measurements with objective refraction (OR) and subjective refraction (SR) in patients with myopia.
The ICC value of central and peripheral refraction were all higher than 0.97 with or without cycloplegia. The peripheral refraction in the nasal, temporal, superior, and inferior quadrants was slightly worse than other parameters, with the largest error interval being 1.43 D. The 95% LoA of the central refraction and OR or SR ranged from -0.89 to 0.88 D and -1.24 to 1.16 D without cycloplegia, respectively, and from -0.80 to 0.42 D and -1.39 to -0.84 D under cycloplegia, respectively.